X-Ray Radiation Statistics towards Quality Assurance in Digital Radiography
IFMBE Proceedings. Vol.38: International Symposium on Biomedical Engineering and Medical Physics (ISBEMP 2012) 2013
Jurijs Dehtjars, Lada Bumbure, Tatjana Kirsanova, Vineta Zemīte

Most of diagnostic radiography today is digital. Therefore there is a great possibility to make an express check of an x-ray beam quality using digital approach. The flux of the x-ray photons detected is governed by the Poisson statistics. The deviation of the radiating technique modes leads from the Poisson statistics. Because the single element of a digital detector matrix produces the signal that is directly proportional to the number of the absorbed photons, the brightness distribution is in use to estimate the statistics of the photons. This is demonstrated to evaluate quality of x-ray beam of the digital x-ray diagnostic equipment.


Atslēgas vārdi
x-ray, radiation, radiography
DOI
10.1007/978-3-642-34197-7_46
Hipersaite
http://link.springer.com/chapter/10.1007%2F978-3-642-34197-7_38?LI=true

Dehtjars, J., Bumbure, L., Kirsanova, T., Zemīte, V. X-Ray Radiation Statistics towards Quality Assurance in Digital Radiography. No: IFMBE Proceedings. Vol.38: International Symposium on Biomedical Engineering and Medical Physics (ISBEMP 2012), Latvija, Riga, 10.-12. oktobris, 2012. Berlin: Springer Berlin Heidelberg, 2013, 145.-148.lpp. ISBN 978-3-642-34196-0. e-ISBN 978-3-642-34197-7. ISSN 1680-0737. e-ISSN 1433-9277. Pieejams: doi:10.1007/978-3-642-34197-7_46