Pilot Clinical Study of Novel Ultrasonic Hydration Monitor for Infants
IFMBE Proceedings. Vol.38: International Symposium on Biomedical Engineering and Medical Physics (ISBEMP-2012) 2013
Aleksejs Tatarinovs, N. Sarvazyan, D. Gardovska, L. Eihvalde, I. Kreicberga

Findings from a pilot clinical study of a novel Infant Hydration Monitor (IHM) are presented. IHM is a portable device measuring ultrasound velocity (USV) through the leg muscles in young infants to monitor changes of their hydration status. Study in newborns revealed that USV in their muscles is lower than in adults, indicating higher water content in the tissue by an average of 12%. Observed cases of short term changes of USV can be explained by hydration changes during applied hydration therapy or water detention regiments.


Atslēgas vārdi
hydration; muscle; pediatrics; Ultrasound velocity
DOI
10.1007/978-3-642-34197-7_10
Hipersaite
http://www.scopus.com/inward/record.url?eid=2-s2.0-84870762916&partnerID=40&md5=ca8bd2292994d6b1e2ebe506caa5ede2

Tatarinovs, A., Sarvazyan, N., Gardovska, D., Eihvalde, L., Kreicberga, I. Pilot Clinical Study of Novel Ultrasonic Hydration Monitor for Infants. No: IFMBE Proceedings. Vol.38: International Symposium on Biomedical Engineering and Medical Physics (ISBEMP-2012), Latvija, Rīga, 10.-12. oktobris, 2012. Berlin: Springer Berlin Heidelberg, 2013, 39.-42.lpp. ISBN 978-3-642-34196-0. e-ISBN 978-3-642-34197-7. ISSN 1680-0737. e-ISSN 1433-9277. Pieejams: doi:10.1007/978-3-642-34197-7_10